Fluke 1503 Insulation Tester
View Product Description
Add to Wishlist
The Fluke 1503 Insulation Testers is compact, rugged, reliable, and easy to use. With their multiple test voltages, they are ideal for many troubleshooting, commissioning, and preventative maintenance applications. Additional features, like the remote probe on these tools save both time and money when performing tests.
Features and benefits:
- Insulation test range: 0.1 MW to 2000 MW (1503 only)
- Insulation test voltages: 500 V, 1000 V
- Save both time and money with Automatic calculation of Polarization Index and Dielectric Absorption Ratio
- Make repetitive tests simple and easy with the 1507’s Compare (Pass/Fail) function (1507 only)
- Repetitive or hard-to-reach testing is easy with the remote test probe
- Live circuit detection prevents insulation test if voltage > 30 V is detected for added user protection
- Auto-discharge of capacitive voltage for added user protection
- AC/DC voltage: 0.1 V to 600 V
- 200 mA Continuity
- Resistance: 0.01 W to 20.00 KW
- Save battery power with auto power off
- Read measurements easily with large, backlit display
- CAT IV 600 V overvoltage category rating for added user protection
- Remote probe, test leads, probes and alligator clips included with each tester
- Accepts optional Fluke TPAK magnetic hanging system to free your hands for other work
- Four AA alkaline batteries (NEDA 15 A or IEC LR6) for at least 1000 insulation tests
- One-year warranty
Included accessories
- TL224 Test Leads,
- TP74 Test Probes
- Clips PN 1958654 (red) and PN 1958646 (black)
- Holster and remote probe
General Specifications
- Maximum voltage applied to any terminal
600 V ac rms or dc - Storage temperature
-40 °C to 60 °C
(-40 °F to 140 °F) - Operating temperature
-20 °C to 55 °C
(-4 °F to 131 °F) - Temperature coefficient
0.05 x (specified accuracy)
per °C for temperatures < 18 °C or > 28 °C
(< 64 °F or > 82 °F) - Relative humidity noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F) - Vibration
Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument - Shock
1 meter drop per IEC 61010-1 2nd Edition
(1 meter drop test, six sides, oak floor) - Electromagnetic compatibility: In an RF field of
3 V/M, accuracy = specified accuracy
(EN 61326-1:1997) - Safety
Complies with ANSI/ISA 82.02.01 (61010-1)
2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN
61010-1 2nd Edition for measurement category IV
600 V (CAT IV) - Certifications
CSA per standard CSA/CAN C22.2
No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition - Batteries
Four AA batteries (NEDA 15A or IEC LR6) - Battery life Insulation test use
Tester can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MW with a duty cycle of 5 seconds on and 25 seconds off. - Battery Life Resistance measurements
Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 W with a duty cycle of 5 seconds on and 25 seconds off. - Size
5.0 cm H x 10.0 cm W x 20.3 cm L
1.97 in H x 3.94 in W x 8.00 in L - Weight
550 g (1.2 lb) - IP rating
IP40 - Altitude Operating
2000 m CAT IV 600 V,
3000 m CAT III 600 V - Non-operating (storage)
12,000 m - Over-range capability
110 % of range
Technical Specification
| Maximum voltage between any terminal and earth ground | 600 V |
|---|---|
| Surge protection | 6 kV peak per IEC 61010-1 600 V CAT III, Pollution Degree 2 |
| Fuse for A input | 11 A, 1000 V FAST Fuse (Fluke PN 803293) |
| Display | Digital: 6,000 counts, updates 4/sec |
| Bar graph | 33 segments, updates 32/sec |
| Operating temperature | -10 °C to + 50 °C |
| Storage temperature | -40 °C to + 60 °C |
| Battery | 9 volt Alkaline, NEDA 1604A/ IEC 6LR61 |
| Battery life | 400 hours typical, without backlight |
Accuracy Specifications
| Measurement | Range | Resolution | Accuracy ± ([% of reading] + [counts]) |
|---|---|---|---|
| DC millivolts | 600.0 mV | 0.1 mV | 2.0 % + 3 |
DC volts |
6.000 V |
0.001 V | |
| 60.00 V | 0.01 V |
||
| 600.0 V | 0.1 V | ||
| Auto volts | 600.0 V |
0.1 V | 2.0 % + 3 (dc, 45 Hz to 500 Hz) |
| 4.0 % + 3 (500 Hz to 1 kHz) | |||
AC millivolts1 True RMS |
600.0 mV |
0.1 mV | 1.0 % + 3 (dc, 45 Hz to 500 Hz) |
2.0 % + 3 (500 Hz to 1 kHz) |
|||
| AC volts1 true-rms | 6.000 V |
0.001 V | 1.0 % + 3 (45 Hz to 500 Hz) 2.0 % + 3 (500 Hz to 1 kHz) |
60.00 V |
0.01 V | ||
| 600.0 V | 0.1 V | ||
| Continuity | 600 O | 1 O | Beeper on < 20 O off > 250 O; detects opens or shorts of 500 µs or longer. |
Ohms |
600.0 O | 0.1Ω | 0.9 % + 2 |
| 6.000 kΩ | 0.001 kΩ |
0.9 % + 1 | |
| 60.00 kΩ | 0.01 kΩ | ||
| 600.0 kΩ | 0.1 kΩ | ||
6.000 MΩ |
0.001 MΩ | ||
| 40.00 MΩ | 0.01 MΩ | 1.5 % + 2 |
|
| Diode test | 2.000 V | 0.001 V | 0.9 % + 2 |
| Capacitance | 1000 nF | 1 nF | 1.9 % + 2 |
10.00 µF |
0.01 µF | ||
| 100.0 µF | 0.1 µF | ||
| 9999 µF | 1 µF | ||
100 µF to 1000 µF |
1.9 % + 2 |
||
| > 1000 µF | 5 % + 20 | ||
| Lo-Z capacitance | 1 nF to 500 µF | 10 % + 2 typical | |
| AC amps true-rms (45 Hz to 500 Hz) | 6.000 A | 0.001 A | 1.5 % + 3 |
10.00 A |
0.01 A | ||
| 20 A continuous overload for 30 seconds max. | |||
DC amps |
6.000 A | 0.001 A | 1.0 % + 3 |
| 10.00 A | 0.01 A | ||
| 20 A continuous overload for 30 seconds max. | |||
| Hz (V or A input)2 | 99.99 Hz |
0.01 Hz |
0.1 % + 2 |
| 999.9 Hz | 0.1 Hz | ||
9.999 kHz |
0.001 kHz | ||
| 50.00 kHz | 0.01 kHz | ||