Fluke 1507 Insulation resistance Tester
Email to a Friend
View Product Description
Add to Wishlist
View Product Description
Add to Wishlist
We supply insulation testers, click to see the latest fluke 1507
The Fluke 1507 Insulation Testers is compact, rugged, reliable, and easy to use. With their multiple test voltages, they are ideal for many troubleshooting, commissioning, and preventative maintenance applications. Additional features, like the remote probe on these tools save both time and money when performing tests.
Features and benefits:
- Insulation test range: 1507: 0.01 MW to 10 GW
- Insulation test voltages: 1507: 50 V, 100 V, 250 V, 500 V, 1000 V (1507 only)
- Save both time and money with Automatic calculation of Polarization Index and Dielectric Absorption Ratio
- Make repetitive tests simple and easy with the 1507’s Compare (Pass/Fail) function (1507 only)
- Repetitive or hard-to-reach testing is easy with the remote test probe
- Live circuit detection prevents insulation test if voltage > 30 V is detected for added user protection
- Auto-discharge of capacitive voltage for added user protection
- AC/DC voltage: 0.1 V to 600 V
- 200 mA Continuity
- Resistance: 0.01 W to 20.00 KW
- Save battery power with auto power off
- Read measurements easily with large, backlit display
- CAT IV 600 V overvoltage category rating for added user protection
- Remote probe, test leads, probes and alligator clips included with each tester
- Accepts optional Fluke TPAK magnetic hanging system to free your hands for other work
- Four AA alkaline batteries (NEDA 15 A or IEC LR6) for at least 1000 insulation tests
- One-year warranty
Included accessories
- TL224 Test Leads,
- TP74 Test Probes
- Clips PN 1958654 (red) and PN 1958646 (black)
- Holster and remote probe
General Specifications
- Maximum voltage applied to any terminal
600 V ac rms or dc - Storage temperature
-40 °C to 60 °C
(-40 °F to 140 °F) - Operating temperature
-20 °C to 55 °C
(-4 °F to 131 °F) - Temperature coefficient
0.05 x (specified accuracy)
per °C for temperatures < 18 °C or > 28 °C
(< 64 °F or > 82 °F) - Relative humidity noncondensing
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F) - Vibration
Random, 2 g, 5-500 Hz per
MIL-PRF-28800F, Class 2 instrument - Shock
1 meter drop per IEC 61010-1 2nd Edition
(1 meter drop test, six sides, oak floor) - Electromagnetic compatibility: In an RF field of
3 V/M, accuracy = specified accuracy
(EN 61326-1:1997) - Safety
Complies with ANSI/ISA 82.02.01 (61010-1)
2004, CAN/CSA-C22.2 NO. 61010-1-04, and IEC/EN
61010-1 2nd Edition for measurement category IV
600 V (CAT IV) - Certifications
CSA per standard CSA/CAN C22.2
No. 61010.1-04; TUV per standard IEC/EN 61010-1 2nd Edition - Batteries
Four AA batteries (NEDA 15A or IEC LR6) - Battery life Insulation test use
Tester can perform at least 1000 insulation tests with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into 1 MW with a duty cycle of 5 seconds on and 25 seconds off. - Battery Life Resistance measurements
Tester can perform at least 2500 earth bond resistance measurements with fresh alkaline batteries at room temperature. These are standard tests of 1 W with a duty cycle of 5 seconds on and 25 seconds off. - Size
5.0 cm H x 10.0 cm W x 20.3 cm L
1.97 in H x 3.94 in W x 8.00 in L - Weight
550 g (1.2 lb) - IP rating
IP40 - Altitude Operating
2000 m CAT IV 600 V,
3000 m CAT III 600 V - Non-operating (storage)
12,000 m - Over-range capability
110 % of range
Technical Specification
AC/DC voltage measurement
Accuracy
| Range | Resolution | 50 Hz to 400 Hz ± (% of Rdg + Digits) |
|---|---|---|
| 600.0 V | 0.1 V | ± (2 % + 3) |
- Input impedance
3 MW (nominal), < 100 pF - Common mode rejection ratio (1 kW unbalanced)
> 60 dB at dc, 50 or 60 Hz - Overload protection
600 V rms or dc
Earth bond resistance measurement
| Range | Resolution | Accuracy1 + (% of Rdg + Digits) |
|---|---|---|
| 20.00 O | 0.01 O | ± (1.5 % + 3) |
| 200.0 O | 0.1 O | |
| 2000 O | 1 O | |
| 20.00 kΩ | 0.01 kΩ |
1Accuracies apply from 0 to 100 % of range.
- Overload protection
2 V rms or dc - Open circuit test voltage
> 4.0 V, < 8 V - Short circuit current
> 200.0 mA
Insulation specifications
- Measurement range
0.01 MW to 2000 MW - Test voltages
50 V, 100 V, 250 V, 500 V, 1000 V - Test voltage accuracy
+ 20 %, - 0 % - Short-circuit test current
1 mA nominal - Auto discharge
Discharge time < 0.5 second for C = 1 μF or less - Live circuit detection
Inhibit test if terminal voltage > 30 V prior to initialization of test - Maximum capacitive load
Operable with up to 1 μF load
Accuracy
| Output Voltage | Display Range | Resolution | Test Current | Accuracy ± (% of Rdg + Digits) |
|---|---|---|---|---|
| 50 V | 0.01 MΩ to 20.00 MΩ | 0.01 MΩ | 1 mA @ 50 kΩ | ± (3 % + 5) |
| (0 % to + 20 %) | 20.0 MΩ to 50.0 MΩ | 0.1 MΩ | ||
| 100 V | 0.01 MΩ to 20.00 MΩ | 0.01 MΩ | 1 mA @ 100 kΩ | ± (3 % + 5) |
| (0 % to + 20 %) | 20.0 MΩ to 100.0 MΩ | 0.1 MΩ | ||
| 250 V | 0.01 MΩ to 20.00 MΩ | 0.01 MΩ | 1 mA @ 250 kΩ | ± (1.5 % + 5) |
| (0 % to + 20 %) | 20.0 MΩ to 200.0 MΩ | 0.1 MΩ | ||
| 500 V (0 % to + 20 %) | 0.01 MΩ to 20.00 MΩ | 0.01 MΩ | 1 mA @ 500 kΩ | ± (1.5 % + 5) |
| 20.0 MΩ to 200.0 MΩ | 0.1 MΩ | |||
| 200 MΩ to 500 MΩ | 1 MΩ | |||
| 1000 V (0 % to + 20 %) | 0.1 MΩ to 200.0 MΩ | 0.1 MΩ | 1 mA @ 1 MΩ | ± (1.5 % + 5) |
| 200 MΩ to 2000 MΩ | 1 MΩ | |||
| 2.0 GO to 10.0 GO | 0.1 GO | ± (10 % + 3) |
EN61557 Specifications
The following tables are a requirement for European labeling.
| Measurement | Intrinsic Uncertainty | Operating Uncertainty1 |
|---|---|---|
| Volts | ± (2.0 % + 3) | 30 % |
| Earth Bond Resistance | ± (1.5 % + 3) | 30 % |
| Insulation Resistance | Depends on test voltage and range. See Insulation Test specifications. | 30 % |
1This specification comes from the standard and indicates the maximum amount allowable by the standard.
EN61557 influence variables and uncertainties
| Earth Bond Resistance | Uncertainty for | Uncertainty for | |
|---|---|---|---|
| Influence Variable | Designation per EN61557 | Insulation Resistance | Earth Bond Resistance |
| Supply Voltage | E2 | 5 % | 5 % |
| Temperature | E3 | 5 % | 5 % |
1Specification confidence level 99 %.
The following tables can be used to determine the maximum or minimum display values considering maximum instrument operating error per EN61557-1, 5.2.4.
Insulation resistance maximum and minimum display values
| 50 V | 100 V | 250 V | 500 V | 1000 V | |||||
|---|---|---|---|---|---|---|---|---|---|
| Minimum | Minimum | Minimum | Minimum | Minimum | |||||
| Limit Value | Display | Limit Value | Display | Limit Value | Display | Limit Value | Display | Limit Value | Display |
| Value | Value | Value | Value | Value | |||||
| 0.05 | 0.07 | 0.05 | 0.07 | 0.05 | 0.07 | 0.05 | 0.07 | ||
| 0.06 | 0.08 | 0.06 | 0.08 | 0.06 | 0.08 | 0.06 | 0.08 | ||
| 0.07 | 0.09 | 0.07 | 0.09 | 0.07 | 0.09 | 0.07 | 0.09 | ||
| 0.08 | 0.10 | 0.08 | 0.10 | 0.08 | 0.10 | 0.08 | 0.10 | ||
| 0.09 | 0.12 | 0.09 | 0.12 | 0.09 | 0.12 | 0.09 | 0.12 | ||
| 0.1 | 0.13 | 0.1 | 0.13 | 0.1 | 0.13 | 0.1 | 0.13 | 0.1 | 0.1 |
| 0.2 | 0.26 | 0.2 | 0.26 | 0.2 | 0.26 | 0.2 | 0.26 | 0.2 | 0.3 |
| 0.3 | 0.39 | 0.3 | 0.39 | 0.3 | 0.39 | 0.3 | 0.39 | 0.3 | 0.4 |
| 0.4 | 0.52 | 0.4 | 0.52 | 0.4 | 0.52 | 0.4 | 0.52 | 0.4 | 0.5 |
| 0.5 | 0.65 | 0.5 | 0.65 | 0.5 | 0.65 | 0.5 | 0.65 | 0.5 | 0.7 |
| 0.6 | 0.78 | 0.6 | 0.78 | 0.6 | 0.78 | 0.6 | 0.78 | 0.6 | 0.8 |
| 0.7 | 0.91 | 0.7 | 0.91 | 0.7 | 0.91 | 0.7 | 0.91 | 0.7 | 0.9 |
| 0.8 | 1.04 | 0.8 | 1.04 | 0.8 | 1.04 | 0.8 | 1.04 | 0.8 | 1.0 |
| 0.9 | 1.17 | 0.9 | 1.17 | 0.9 | 1.17 | 0.9 | 1.17 | 0.9 | 1.2 |
| 1.0 | 1.30 | 1.0 | 1.30 | 1.0 | 1.30 | 1.0 | 1.30 | 1.0 | 1.3 |
| 2.0 | 2.60 | 2.0 | 2.60 | 2.0 | 2.60 | 2.0 | 2.60 | 2.0 | 2.6 |
| 3.0 | 3.90 | 3.0 | 3.90 | 3.0 | 3.90 | 3.0 | 3.90 | 3.0 | 3.9 |
| 4.0 | 5.20 | 4.0 | 5.20 | 4.0 | 5.20 | 4.0 | 5.20 | 4.0 | 5.2 |
| 5.0 | 6.50 | 5.0 | 6.50 | 5.0 | 6.50 | 5.0 | 6.50 | 5.0 | 6.5 |
| 6.0 | 7.80 | 6.0 | 7.80 | 6.0 | 7.80 | 6.0 | 7.80 | 6.0 | 7.8 |
| 7.0 | 9.10 | 7.0 | 9.10 | 7.0 | 9.10 | 7.0 | 9.10 | 7.0 | 9.1 |
| 8.0 | 10.40 | 8.0 | 10.40 | 8.0 | 10.40 | 8.0 | 10.40 | 8.0 | 10.4 |
| 9.0 | 11.70 | 9.0 | 11.70 | 9.0 | 11.70 | 9.0 | 11.70 | 9.0 | 11.7 |
| 10.0 | 13.0 | 10.0 | 13.0 | 10.0 | 13.0 | 10.0 | 13.0 | 10.0 | 13.0 |
| 20.0 | 26.0 | 20.0 | 26.0 | 20.0 | 26.0 | 20.0 | 26.0 | 20.0 | 26.0 |
| 30.0 | 39.0 | 30.0 | 39.0 | 30.0 | 39.0 | 30.0 | 39.0 | 30.0 | 39.0 |
| 40.0 | 52.0 | 40.0 | 52.0 | 40.0 | 52.0 | 40.0 | 52.0 | 40.0 | 53.0 |
| 50.0 | 65.0 | 50.0 | 65.0 | 50.0 | 65.0 | 50.0 | 65.0 | ||
| 60.0 | 78.0 | 60.0 | 78.0 | 60.0 | 78.0 | 60.0 | 78.0 | ||
| 70.0 | 91.0 | 70.0 | 91.0 | 70.0 | 91.0 | 70.0 | 91.0 | ||
| 80.0 | 104.0 | 80.0 | 104.0 | 80.0 | 104.0 | 80.0 | 104.0 | ||
| 90.0 | 117.0 | 90.0 | 117.0 | 90.0 | 117.0 | 90.0 | 117.0 | ||
| 100.0 | 130.0 | 100.0 | 130.0 | 100.0 | 130.0 | ||||
| 200.0 | 260.0 | 200.0 | 260.0 | ||||||
| 300.0 | 390.0 | 300.0 | 390.0 | ||||||
| 400.0 | 520.0 | 400.0 | 520.0 | ||||||
| 500.0 | 650.0 | ||||||||
| 600.0 | 780.0 | ||||||||
| 700.0 | 910.0 | ||||||||
| 800.0 | 1040.0 | ||||||||
| 900.0 | 1170.0 | ||||||||
| 1000.0 | 1300.0 | ||||||||
| 2000.0 | 2600.0 | ||||||||
Earth bond resistance maximum display values
| Limit Value | Maximum Display Value |
|---|---|
| 0.4 | 0.28 |
| 0.5 | 0.35 |
| 0.6 | 0.42 |
| 0.7 | 0.49 |
| 0.8 | 0.56 |
| 0.9 | 0.63 |
| 1.0 | 0.7 |
| 2.0 | 1.4 |
| 3.0 | 2.1 |
| 4.0 | 2.8 |
| 5.0 | 3.5 |
| 6.0 | 4.2 |
| 7.0 | 4.9 |
| 8.0 | 5.6 |
| 9.0 | 6.3 |
| 10.0 | 7.0 |
| 20.0 | 14.0 |
| 30.0 | 21.0 |
| 40.0 | 28.0 |
| 50.0 | 35.0 |
| 60.0 | 42.0 |
| 70.0 | 49.0 |
| 80.0 | 56.0 |
| 90.0 | 63.0 |
| 100.0 | 70.0 |
| 200.0 | 140.0 |
| 300.0 | 210.0 |
| 400.0 | 280.0 |
| 500.0 | 350.0 |
| 600.0 | 420.0 |
| 700.0 | 490.0 |
| 800.0 | 560.0 |
| 900.0 | 630.0 |
| 1000.0 | 700.0 |
| 2000.0 | 1400.0 |